USER MANUAL
SOLDERABILITY TESTER
METRONELEC
V2.0
–
10/2021
6 Avenue Eiffel - 78420 Carrières-sur-Seine
Tél : 01 30 15 20 00
Web :
http://www.metronelec.com| Mail : [email protected]
Page
64
sur
83
TYPE 19 : - PC BOARD OR COMPONENT PARTIALLY WETTABLE
APPLICATION FIELD: PCB with stripes on one side or two sides. Complex or specific
component.
PCB stripe width or wettable perimeter
40% sample perimeter.
DIMENSIONS in mm
TOOL ref. : D19 Put the sample in the tool holder, let only few mm of component
termination for a conventional component (solder wave) or less for a SMD component.
PARAMETERS:
TIME:
Select between 5 - 10 - 15 - 20 seconds to obtain the stabilized force before end of
measurement
IMMERSION DEPTH:
0.5 or 1 mm
SENSITIVITY:
Adjustment depends of the component size. The sensitivity is automatically selected by
the software.
FLUX:
Dip the extremity of sample (epoxy and metal) in the flux, and remove the drop on the
absorbent paper.