Display Start Windows
: Display or not the application selection windows
, where we
choose between, GammaLV, flux efficiency, component solderability, etc
Crossing time
: Crossing time is the time when the curve cut the buoyancy level (Ta) or
the “X” axis level (Tb). Both are used in different standards but on the graph only one is
calculated so you can select buoyancy or “zero” crossing time.
Example curve of “buoyancy
crossing time”
Crossing time is 0.26 s between the curve and the buoyancy level (discontinuous pink line)