USER MANUAL
SOLDERABILITY TESTER
METRONELEC
V2.0
–
10/2021
6 Avenue Eiffel - 78420 Carrières-sur-Seine
Tél : 01 30 15 20 00
Web :
http://www.metronelec.com| Mail : [email protected]
Page
24
sur
83
TYPE
Choose the type of component as defined in Section 4.
This choice will set the formulas and the pass fail criteria to use following the shape / type of
component.
Click on the
button to have the list of type to select as below :
ALLOY
Type the name of the alloy or select it in the list by clicking on the lift button.
Alloy choice must be done in a standard list or the closest to the alloy used in production.
It is important to know well the GammaLV value for any couple of Alloy / flux and the alloy
density at the test temperature as it is used to define the maximum force and the wetting
angle