You can approach vertically the globule from the sample with the fast up arrow, when
aligned, just press enter in the previous window.
10.
The inert chamber system goes down (if requested).
11.
The receptacle goes up until the desired immersion depth
12.
Measurement
13.
Receptacle, inert chamber come back to home position
14.
Remove the tool holder and the component from the sensor
15.
Clean the tool holder using solvent
16.
Software asks for data saving and for next measurement with same parameters.
Analysis of results
Display the curve
Once the measurements have been taken, the results are displayed in a graph.
Each measurement has a number (increasing order), regardless of whether it is validated or
not, this appears at the top of the graph and on the browser in the right windows (if
opened!).