83
Fmin) / Fmax. Standard accept generally a value of 20 %. Don’t let a value of 0 as we ha
ve
often a dewetting of few percent…if not the test will be failed.
Comment
: space for operator to describe the conditions of test, batch number, etc
Example of dewetting curve Fmax Fmin.
Other Spec (specification)
In order to create personal specification
Name
: Select the user standard wanted to apply on graph. Or type a new one to save a
new standard.
Time
: T1 and T2 times for wetting angle Theta1 and theta 2 and for force.
Fmin
Fmax