MI 2087 Instaltest 61557
Memory and other operations
55
All test results can be stored, except the following: Continuity (auxiliary function),
Phase rotation (comparative test) and Voltage/Frequency.
How to store displayed test result?
Result is thus stored, note used FUSE CABINET and FUSE codes down to electric
diagram of tested installation for later recognition of stored result.
Main result, potential subresults and potential function parameters as well as Function
identification number (see the figure below) are stored, whenever “Store” procedure is
affected.
More test results (regardless of test function) can be stored under the same FUSE
CABINET/FUSE code. This will be done simply, effecting double click to
SAVE
key.
FUL
message will be displayed in case all memory locations are occupied.
LOCATOR
U
C
L
,
R
R
LINE SC
,
I
R
LOOP SC
,
I
U
L-N
U
L-PE
t
D
N
U
L-N
LOG
I
D D
,
t
RCD
AUTO
9
8
7
6
10
11
12
1
2
3
R
±200mA
4
5
SENSOR
Function identification
numbers
Fig. 52.
Function identification numbers
Aim of storing the identification number is to make clear which function recalled result
belongs to.
FUSE code is offered (blinking).
Set FUSE code (001 up to 999 can be set).
FUSE CABINET code is offered (blinking).
Set FUSE CABINET code (001 up to 999 can be set).
Store displayed result under set FUSE / FUSE CABINET code.