Measurement instructions
49
3.13. Illumination
How to carry out the illumination measurement?
Step 1
•
Connect LUXmeter probe to Instaltest 61557.
•
Set function switch to
SENSOR
position, the following menu will be displayed.
Fig. 43.
Illumination measurement initial menu
Step 2
•
Set
Low limit Illumination value
. Later will measured results be compared with
the set limit value and, if lower, they will be equipped with “!” mark.
How to set the Low limit value?
•
Press the
SELECT
key, to enter “Low Limit value adjustment mode”, the
following menu will appear:
Fig. 44.
Low limit value adjustment menu and the table of available limit values
•
Use the
↑
and
↓
keys, to set required Low Limit Value. Individual strike will
increase/decrease the value for one step, while continuous pressure will
increase/decrease it continuously. If test results are not to be compared with
Low Limit Value at all, then press the
CLR
key,
no
will be displayed instead of
set value.