MI 3155
Eurotest
XD
Test and measurements
167
7.29
Illumination
Figure 7.99: Illumination measurement menu
Measurement parameters / limits
Limit(E) Minimum illumination
[Off, 0.1 lux ... 20 klux]
Probe positioning
Figure 7.100: LUXmeter probe positioning
Measurement procedure
Enter the
Illumination
function.
Set test parameters / limits.
Connect illumination sensor A 1172 or A 1173 to the instrument.
Take the position of LUXmeter probe, see
Make sure that LUXmeter probe is turned on.
Start the continuous measurement.
Stop the measurement.
Save results (optional).