MI 3155
Eurotest
XD
Appendix F
– Programming of AutoSequences
on Metrel ES Manager
237
Operation after end of test
This flow command controls the proceeding of the Auto Sequence
® in regard to the
measurement results.
Parameters
Operation after end of test
pass
fail
no status
The operation can be individually set for the case the
measurement passed, failed or ended without a status.
Manual:
The test sequence stops and waits for
appropriate command (RUN key, external
command
…) to proceed.
Auto:
The test sequence automatically proceeds.
Result screen
This flow command controls the proceeding after the Auto Sequence
® has ended.
Parameters
Auto save
Auto
Sequence® results are stored in the momentary workspace.
A new Node with the current month and year will be created. Under the
Node Auto
Sequence® results will be stored.
Up to 100 Auto
Sequence® results can be automatically stored under the
same node. If more results are available, they are split to multiple nodes.
Auto Save Flow setting is disabled by default.
Note
This flow command is active only if Auto
Sequence® is started from
the Auto Sequence® Main menu (not from the Memory organizer).