User Guide Version 2.60 I-Button
Revision 21_Sep_2015
13
NOTE:
All Set-up fields
must have data in
order to transfer
information to the
SQA-V. If CONTROL
settings are not known,
enter “0” LOT #/
Target Value/+/-
Range. Enter current
date for the date field.
NOTE
: The
Set-up
data transfer may take
several minutes!
Please wait…..
NOTE:
Factory default
settings are listed in
RED
.
SQA-V System Default settings:
Date Format (DD/MM/YY) or (MM/DD/YY)
Local date setting
Conc./Chamber Standard 1 or 2 (See appendix section for more information).
LES setting: Check with your distributor for set-up
Printing options: automatically print test results/self test report on start-up.
Control Set-up (from the manufacturer’s labeling):
Select type of control: Latex beads or Stabilized Sperm CAP.
Enter Lot Number for each control level (enter “0” if not known).
Enter +/- Range for each control level (enter “0” if not known).
Enter EXPIRATION date (use current date if the EXP date is unknown).
Press the
Report
button to view and print the selected default settings.
Press
Apply
to accept the default settings and transfer them to the SQA-V.
Testing
Samples
SECTION 5: Testing Semen Samples
Information about the patient and sample is entered prior to the testing process. In
order to accurately "classify" the semen sample by type and volume and understand
the options for testing, refer to the information below.
Patient
Information
Entering Patient and Sample Information
ENTER PATIENT / SAMPLE DATA
PATIENT ID:
5788114
BIRTH DATE:
01/01/85
ABSTINENCE:
4 DAYS
SAMPLE PROCESSING
SAMPLE / ACCESSION # 88
COLLECTED: DD/MM/YY HH:MM
RECEIVED: DD/MM/YY HH:MM
From the
MAIN MENU
select
TEST NEW PATIENT
and the
ENTER PATIENT/
SAMPLE DATA
screen is displayed.
Enter the requested sample/patient information using the SQA-V keypad:
PATIENT ID
– Unique number identifying the patient (Maximum of 20
numbers can be entered).
BIRTH DATE
– Birth date of the patient.
ABSTINENCE
- Number of days since the patient's last ejaculation.
SAMPLE/ACCESSION #
- Up to 20 numbers identifying the sample.
COLLECTED –
Date and time the sample was collected.
RECEIVED –
Date and time the sample was received.
PLEASE NOTE:
The SQA-V is calibrated
to run semen
specimens at room
temperature. It is not
necessary nor will the
user get accurate
motility results if the
sample is heated to
37ºC.