User Guide Version 2.60 I-button WHO 5th
Revision 21_Sep_2015
12
SQA-V set-up
screen from
V-Sperm GOLD
CONTROL set-up
screen from
V-Sperm GOLD
Set-up System Defaults
SQA-V system defaults are set-up through V-Sperm GOLD software. Therefore a
connection needs to be established between the SQA-V and the PC.
From the
MAIN MENU
, select
SERVICE > SERVICE DATA
.
SERVICE MENU
SERVICE DATA
SERVICE PERSONNEL
PRINT SELF-TEST DATA
ADD I-BUTTON TESTS
SERVICE DATA
1. 18
8. 112 15.
1.3
2. 5 9. 10 16.
110
3.
150 10. 6 17. 2
4. 28
11. 89 18. 1000
5. 77.65
12. 31
19. 1
6. 512
13. 100
7. 0.000
14. 100
The RS232 communication cable must be connected to the SQA-V and the PC.
Turn-on the PC and activate the V-Sperm GOLD version 3.60 software.
From the V-Sperm GOLD main navigation screen select
SET-UP
>
SQA-V >
SQA-V Defaults
. Then press the
CONTINUE
button.
V-Sperm GOLD will display the SQA-V system set-up screen: