50
06/2021
Cooling unit CDP
7.2.5 Setting up
WARNING
The tube is supplied in a predefined length. This is possibly too long. Please make
sure that the hoses do not hang down. You could get caught during operation and
suffer scalds or burns from the heated water.
CAUTION
The appliance may fall from the installation location due to unevenness and injure you
or someone else. Always place the appliance on an even, load-bearing and non-slip
surface.
CAUTION
The surfaces of the appliance can overheat during operation and overheat heat-sen-
sitive surfaces in close proximity to the appliance. Always place the appliance on a
heat-resistant surface and maintain the specified clearance around the appliance.
The distance between the wall and the rear of the appliance must be at least 15 cm. The
clearance from the ceiling must not be less than 75 cm and the side clearance from
walls or nearby appliances must not be less than 10 cm. Sufficient air circulation in the
vicinity of the appliance must be guaranteed at all times.
NOTICE
Determine the installation location before you start installing the cooling unit to the water bath. The water bath and cooling unit may
only be transported individually.
10 cm
10 cm
15 cm
10 cm
75 cm