41
06/2021
Pumping device CPM
6.4
Operation and control
WARNING
During operation of the pumping device, individual components may heat up.
Touching its surface can cause burns. Use temperature-resistant protective gloves.
6.4.1 Switching on appliance
Press main switch on the back of the appliance.
The pumping capacity of the pumping device is constant and cannot be regulated.
While switched on, the pumping device operates at a steady rate. The temperature is
controlled via the setting on the water bath.
6.4.2 Function test
■
To check that the unit is pumping, pull off the suction tube from the tube suspender and
place a finger on it to test for suction. If there is no suction effect: See chapter Faults,
warning and error messages.
6.5
Dismantling the hoses
WARNING
During operation of the pumping device, individual components may heat up.
Touching its surface can cause burns. Let the heated components cool down first.
1.
Allow the wa
ter in the water bath and the hoses to cool down first.
2.
Drain the water from the water bath using the drain tap
until the raw joints are exposed.
3.
Pull the hoses off the pipe connections
NOTICE