51
06/2021
Cooling unit CDP
7.2.6 Installation cooling unit
1.
Attach brackets for the pipe connection to the back of the water bath.
2.
Attach the pipe connections to the brackets and hang them in the water bath.
NOTICE
Pump connection IN (suction line): Attach the tube to the long pipe connec-
tion on the water bath, as this is where the water is conveyed from the water
bath to the pumping device.
Pump connection OUT (pressure line): Attach the tube to the short pipe
connection on the water bath, as this is where the water from the pump is
conveyed back to the water bath.