Cooling unit CDP
7.1.10 Ambient conditions
Ambient temperature
+5 °C to 40 °C
Air humidity
max. 80 % rh (non-condensing)
Overvoltage category
II
Pollution degree
2
Altitude of installation
max. 2,000 m above sea level
Maximum mains voltage fluctuations
AC 115 (+/– 10 %)
AC 230 (+/– 10 %)
■
The appliance may only be used in enclosed areas and under the ambient
conditions listed below
■
The appliance may not be used in areas where there is a risk of explosion. The sur-
rounding air must not contain any explosive dusts, gases, vapours or gas-air mixtures.
The appliance is not explosion-proof.
■
Heavy dust production or aggressive vapours in the vicinity of the appliance could lead
to sedimentation in the interior and, as a consequence, could result in short circuits or
damage to electrical parts. For this reason, sufficient measures to prevent large clouds
of dust or aggressive vapours from developing should be taken.
7.1.11 Scope of delivery and required accessories
■
Cooling unit
■
Installation and connection set (consisting of inlet and outlet pipe, two hinges)