SAFETY WARNINGS

Safety warnings and precautions must be read and 

understood before the instrument is used. They must be 
observed during use.

The circuit under test must be switched  off, de -
energised and isolated before Insulation or Continuity 
tests are made.

The test button must not be held down while connecting 
the test leads or while changing ranges. (May cause 
Live Circuit Warning‘ to become inoperable).

The Voltage warning does not function if OFF or 
is selected.

During an insulation test, connections must not be 
touched.

After insulation tests, capacitive circuits must be allowed 
to discharge before disconnecting the test leads.

Test leads, prods and crocodile clips must be in good 
order; clean, and with no broken or cracked insulation.

Replacement fuses must be of the correct size, type and 
rating.

NOTE

THE INSTRUMENTS MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT

PERSONS

.

Symbols used on the instrument:

Caution: risk of electric shock

Caution: refer to accompanying notes

Equipment protected throughout by Double Insulation (Class 

II

)

Equipment complies with relevant EU Directives

Summary of Contents for BM220 Series

Page 1: ...to become inoperable The Voltage warning does not function if OFF or is selected During an insulation test connections must not be touched After insulation tests capacitive circuits must be allowed to...

Page 2: ...strument On by selecting the range 2 If required zero test lead resistance by firmly shorting both leads together wait for the reading to stabilise and press the test button Display of the symbol conf...

Page 3: ...Voltage 5V 1V Accuracy at 20 C 0 01 9 99 3 2 digits ISC 200mA 10 100 5 2 digits ISC 20mA Lead res comp 0 to 9 99 Hum Rejection 3 error with 1V RMS 0 2 to 50 Continuity Beeper On position the beeper s...

Page 4: ...batteries Fuse 500 mA F H B C 10 kA min 32mm x 6mm To check this fuse select M open circuit the test leads and press the test button until a reading is obtained Display of the fuse symbol or an error...

Page 5: ...roved Repair Companies A number of independent instrument repair companies have been authorised for repair work on most Megger Instruments using genuine Megger spare parts Consult the Appointed Distri...

Page 6: ...038 PO Box 9007 Valley Forge PA 19484 9007 Tel 1 610 676 8500 U S A Fax 1 610 676 8610 Megger SARL 29 All e de Villemomble 93340 Le Raincy Paris Tel 33 1 43 02 37 54 France Fax 33 1 43 02 16 24 This i...

Page 7: ...le confirme la mise z ro Nota La mise z ro des c bles est annul e chaque fois que l instrument est arr t ou que la fonction Arr t automatique est activ e 3 Brancher les sondes d essai sur le circuit i...

Page 8: ...3 c c non sp cifi D charge automatique Les circuits capacitifs sont automatiquement d charg s lorsque le bouton d essai est rel ch apr s un essai d isolement Alimentation lectrique 6 piles 1 5V type I...

Page 9: ...i jedem Ausschalten des Ger ts oder Aktivieren der automatische Abschaltfunktion wieder aufgehoben 3 Pr fspitzen an zu pr fenden Stromkreis anschlie en 4 Auf der Anzeige wird der Widerstand angezeigt...

Page 10: ...C keine Angaben Automatische Entladung Beim Loslassen der Pr ftaste nach einer Isolierpr fung werden die kapazitiven Stromkreise automatisch entladen Stromversorgung sechs 1 5 V Zellen nur IEC LR6 Leb...

Page 11: ...puesta a cero del conductor se cancela cada vez que se desconecta el instrumento o se activa desconexi n autom tica 3 Conecte las sondas de prueba al circuito aislado en prueba 4 El display muestra el...

Page 12: ...apacitivos son descargados autom ticament cuando se suelta el bot n de prueba despu s de una prueba de aislamiento Suministro de energ a 6 pilas de 1 5V de tipo IEC RL6 solamente Vida til de la bater...

Reviews: