9
Alarm
Codes
Fault Type
Description
Troubleshooting
LO
Low Pressure
• Leg is less than 40mmHg
pressure; foot pressure is
less than 120mmHg after
2 consecutive cycles.
• Air hose disconnected
at sleeve.
• Check sleeve application
to ensure proper fit.
• Check for leaks in the sleeves
or air hose set and replace if
faulty.
• Turn the pump off and restart.
• Refer to a service professional
if problem persists.
HI
High Pressure
• Air hose is kinked.
• Check air hose set for
any kinks.
• Refer to a service professional
if problem persists.
GE
Sleeve Error
• No sleeves attached
during start up.
• Sleeves are not fully
deflated during each restart.
• The initialization process
has detected an inconsis-
tent sleeve configuration.
• The initialization failed due
to a possible blockage in air
hose/sleeve connectors.
• Attach sleeves to pump
and restart.
• Ensure sleeves are fully
deflated before restart.
• Check sleeve configuration
and restart.
• Check connectors for any
blockage and restart.
• Refer to a professional for
service if problem persists.
F
System Failure
• Pressure is greater
than set pressure after
two inflations.
• Electrical malfunction.
• Check air hose connection.
• Turn the pump off and restart.
• Refer to a service professional
if problem persists.
Battery Backup System
• This device has a Lithium Ion battery backup system. The device constantly
monitors the outer AC electric power supply. In the case of a power failure the
pump will switch to battery operation automatically.
• The “Battery Indicator” indicates the battery capacity status as illustrated.
• Under normal conditions a fully charged battery will last approximately 4 hours of
continuous operation. It may take up to 16 hours to fully charge a depleted battery.
70-99%
50-70%
30-50%
5-30%
5%