27
Pressure Adjustments and Tests
1. Make sure the power source is 120 Volt 60 Hz. Use a power regulator
to stabilize the power voltage in order to increase accuracy of pressure setting.
2. Attach the test kit to pump and manometer.
3. Press the Power Micro Switch.
4. Perform the pressure test procedure from Minimum Pressure (LEG)
to Maximum Pressure (FT).
Minimum Pressure Test
1. Press Mode Membrane Button (CALF/THIGH SLEEVES) and LED Situation
Monitor will display “LEG”.
2. Check the pressure is 40 mmHg ( +10/-5 mmHg ).
Maximum Pressure Test
1. Press Mode Membrane Button (FOOT) and LED Situation Monitor will display ”Ft”
2. Check the pressure is 120 mmHg ( +10/-5 mmHg ).
Pressure Adjustment Procedure
1. Open the case and identify pressure regulators.
2. Pressure adjustment can be adjusted with a Philips screwdriver.
Pressure can be adjusted by clockwise/anti-clockwise turns to
decrease/increase the pressure respectively.