4. Application of the calf and thigh sleeve(s).
a. Sleeves come in identical pairs and may be used on either leg.
b. When using dual sleeves the identical sleeve type (same model) must be used
on both legs. Sleeve combinations are calf and calf, thigh and thigh or foot and
foot. Do not use in any other combination.
c. Unfold the sleeve and lay the unfolded sleeve on a flat surface with the
foam lining facing up.
d. Position the sleeve at the appropriate ankle or knee landmark.
e. Place the patient’s leg on the sleeve lining, position the back of the knee
over the popliteal opening in the case of thigh high sleeves.
f. Fasten the sleeve securely with the Velcro® fasteners starting at the ankle
and then moving to the calf and/or thigh. Secure with tabs starting at the
bottom tab marked 1, moving upwards to tab 2 then 3, etc. Check that none of
the Velcro flaps hang off the end of the sleeve (entirety of Velcro area is secured
to the sleeve). The sleeve should fit securely but not tightly around all parts of the
patient’s leg. Confirm sleeve is properly sized by checking that the index line falls
between the range markers. Follow the instructions and select the next size up or
down if the index line falls out of range.
g. Use the “two-finger” rule to check that the correct fit is used. Insert 2 fingers
horizontally (with palm up and both fingers touching the patient’s leg) in between
the sleeve and the patient. This practice should be done at both the top and
bottom of the sleeve.
h. Using the same procedures apply the second identical model sleeve if needed.
If only one sleeve is to be used, simply leave the unused air outlet on the pump
free (no air hose attached).
5. Application of the foot sleeve(s).
a. Foot sleeves come in identical pairs and may be used on either foot.
b. When using dual foot sleeves the identical sleeve type (same model) must
be used on both legs. Do not use in any other combination. Place the sleeve
on the patient’s foot ensuring the Velcro attachments are visible.
c. Ensure a snug fit, by placing two fingers around the back of the sleeve to
ensure the sleeve is not too tight. Routinely inspect the patient’s feet once
compression commences.
d. Using the same procedures, apply the second identical model sleeve if needed.
If only one sleeve is to be used simply leave the unused air outlet on the pump
free (no air hose attached).
6. Air hose(s) are required to connect the sleeve(s) to the pump and are provided
separately. Attach the sleeve(s) to the air hose(s) using the white snap lock
connectors. Each tube has a male end connector at one end and a female
connector at the other. The female end (large white connector) will fit to the male