8

12. To activate the timer, turn the pump on and timing will commence.  
The timer reflects each hour of compression. When compression is interrupted  
i.e. the patient is removed from therapy, the timer will suspend timing until the  
patient is reconnected to therapy and therapy is continued.

13. To reset the timer press “RESET” for 5 seconds and it will reset itself.

14. To turn pump off, hold the power button for 5 seconds.
15. To disconnect device remove plug from electrical outlet.

16. Be sure battery is installed in the back side of the device to avoid 

 

an interruption in therapy due to power failure.

17. Respond to the instrument alarms appropriately and refer to troubleshooting 

guide on pump (page 28 of this manual).
18. To remove the sleeves, press the tab on large white (female) connector 
and pull apart. 

Recommendations during operation
• Avoid placing the tubing and/or any connections or components under the patient 
as this may negatively impact the effectiveness of the pump and/or cause a skin 

integrity issue for the patient

• Make sure that there are no kinks in the tubing and that the connectors are 

properly locked.  

• Never apply or remove the sleeves while inflated as this may cause damage 

to the sleeve.

• The parts and/or accessories supplied are specifically designed for use with 
Hemo-Force IPCS control unit. Use of other products in conjunction with the 

system is not recommended.

Alarms and indicators

Indicator Light: Green

Illuminated to indicate power on

Indicator Light: Amber

Flashing to indicate a fault has been detected follow by 
an audio alarm

Indicator Light: Blue

Flashing indicates initialization in process for 10 sec and 
upper display window showing (Int) illuminated indicates 
unit is operational

Blue LED Display

Upper display indicates therapy selected for Calf/Thigh 
(LEG) or FOOT (Ft). Lower display indicates pressure 

delivered to each chamber of compression and alarm 

status for low pressure (LO), high pressure (HI), sleeve 
error (GE) and fault (F).

Timer

Indication of therapy time in hours from commencement 
of therapy per patient

Summary of Contents for Hemo-Force MDS600INT

Page 1: ...1...

Page 2: ...5 Control Panel Features 5 Operating Instructions 5 Alarms and Indicators 8 Cleaning 10 Electromagnetic Compatibility 11 Technical Specifications 14 Servicing and Maintenance 16 General Disassembly an...

Page 3: ...e the state corresponding to a specified initial number Battery indicator that can show remaining battery status Unique hanging bracket that provides easy attachment Intended Use The system consists o...

Page 4: ...ith hospital policy 5 Keep the pump and sleeve away from sources of liquid and open flames 6 The control unit should only be repaired by an authorized professional 7 Do not drop the control unit or st...

Page 5: ...Flammability as defined in CPSC 16 CFR Part 1610 Control Panel Features Alarm Light Blue LED Display Window Battery Indicator Time in Hours Foot Mode Pressure Settings for Foot at 120mmHg Time Reset...

Page 6: ...two finger rule to check that the correct fit is used Insert 2 fingers horizontally with palm up and both fingers touching the patient s leg in between the sleeve and the patient This practice should...

Page 7: ...d hold the power button to turn pump on when air hose s and sleeve s are properly connected 9 Pump will start initialization process lasting approximately 10 seconds 10 The pump will automatically det...

Page 8: ...may negatively impact the effectiveness of the pump and or cause a skin integrity issue for the patient Make sure that there are no kinks in the tubing and that the connectors are properly locked Neve...

Page 9: ...air hose sleeve connectors Attach sleeves to pump and restart Ensure sleeves are fully deflated before restart Check sleeve configuration and restart Check connectors for any blockage and restart Refe...

Page 10: ...ly before cleaning and inspection The Sleeve It is a single patient use sleeve Do not wash or clean discard after use Caution Do not use phenol based cleaning solutions Switch off the electrical suppl...

Page 11: ...ce d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitte...

Page 12: ...1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power qual ity should be that of a typical commercial or hospital environment interruptions and voltage vari...

Page 13: ...ransmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the fol...

Page 14: ...s Inflation Time 8 Seconds Hold Time 4 Seconds Deflation Time 18 Seconds Display Type LED Alarm Type Audio and Visible Hook attachment Yes Owners Manual Included Battery Pack Model optional MDS600BAT...

Page 15: ...15 Symbols Used Type BF Applied Part Manufacture Date of Manufacture Batch Code Ex YYMMDD01XX Serial Number Caution...

Page 16: ...ain Relief Bushing 6 Control Panel 6 1 Power Micro Switch ON OFF 6 2 Mode Membrane But ton Foot 6 3 Membrane Button Reset for timer reset 7 Function Status LED Light Indication 7 1 Power Light Green 7...

Page 17: ...ery Pack recharge discharge 16hr 4hr Optional equipment C Initial Assessment Procedure Item Tasks 1 Check the Plastic Chassis Housing appearance 2 Check the Mount Fasteners 3 Check the Hooks 4 Check t...

Page 18: ...run for a minimum of 1 hour and check calibration Adjust if required Operational Maintenance Routine checks are recommended to be carried out for each time the system is in use For the control unit 1...

Page 19: ...ter 1 Identify the air filter module on the back the device 2 Press the side lock to release the cap 3 Clean the cap with a soft brush 4 Remove the filter and clean the bottom with a soft brush 5 Inse...

Page 20: ...e back of the device 2 Unscrew the screw to release the cable cover 3 Press the side lock to release the battery pack 4 Remove the battery pack and check the socket for dirt and or debris 5 Inspect th...

Page 21: ...Pack Placement procedure 1 4 to remove the battery pack 3 Unscrew the 6 pan head screws 4 Open the case and keep the screws in a safe place 5 Disconnect 2 air tubes from the valve 6 The control unit...

Page 22: ...sor and valve 4 Unscrew and remove the 2 screws securing the compressor 5 Remove the whole compressor set out of the case 6 Re assemble new components in reverse order Solenoid valve set replacement 1...

Page 23: ...23 2 Unplug the valve power cable from PCB board 3 Disconnect the air tube from the valve 4 Unscrew and remove the 2 screws 5 Remove the whole valve set 6 Re assemble new components in reverse order...

Page 24: ...24 Fuse replacement 1 Identify the fuse module 2 Remove the fuse s 3 Insert the new fuse s 22 23 17 16 21 19 18 20 Schematics...

Page 25: ...PCB Set 10 MDSCH40020 120V Power Cord 11 MDSCML20045 Tube Set For Compressor 12 MDSCML20046 Pressure Tube Set 13 MDSCML20047 Air Tube Set 14 MDSBF004001 Rubber Foot 15 MDSAL200208 Label Front Panel Fo...

Page 26: ...ssure at 120 mmHg Exclusive to Hemo Force IPCS Battery Indicator Time in Hours Time Reset Button Power Button Power Light Single Sleeve Mode Calf Thigh Modede Foot Mode Alarm Light Pressure Display Wi...

Page 27: ...Minimum Pressure Test 1 Press Mode Membrane Button CALF THIGH SLEEVES and LED Situation Monitor will display LEG 2 Check the pressure is 40 mmHg 10 5 mmHg Maximum Pressure Test 1 Press Mode Membrane B...

Page 28: ...obligation under this warranty is limited to the repair of equipment returned to the place of purchase within 12 months of delivery date We agree to service adjust any equipment returned and to replac...