17
B) Initial Assessment (Quantitative Check)
Item
Qualitative
Tasks
Pass
Fail
Service
1
Ground
Resistance less
than < 0.2 ohm
2
Max Leakage
Current less than
< 0.5 mA
3
Pressure Test
3.1
Minimum Pressure
(LEG) 40 +10/ -5
mmHg
3.2
Maximum
Pressure (FT)
120 +10/ -5 mmHg
4
Flow-rate (FT) 6
L/min
5
Battery Pack
(recharge /
discharge)
16hr / 4hr*
*Optional equipment
C) Initial Assessment Procedure
Item
Tasks
1
Check the Plastic Chassis / Housing appearance
2
Check the Mount / Fasteners
3
Check the Hooks
4
Check the AC Plug / Receptacles Outlet
5
Check the power supply unit (Including: Fuses
and Strain Relief Bushing)
6
Press the Power Micro Switch, the Power Light
should illuminate and pump is activated.
7
Follow the alarm troubleshooting guide on the
medical device and remove the basic problems.