Electromagnetic Compatibility
Electric devices may interact due to electro-magnetic radiation. We recommend
a safety distance of at least 3.3 feet (1 meter) especially for sensitive equipment.
Upon request, we will provide you with a table for more detailed information.
Recommended separation distances between portable and mobile
RF communications equipment and the Hemo-Force IPCS
The
Hemo-Force IPCS
is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the
Hemo-Force IPCS
can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the
Hemo-Force IPCS
as recommended below, ac-
cording to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1,2
80 MHz to 800 MHz
d = 1,2
800 MHz to 2,5 GHz
d = 2,3
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Article I. Guidance and manufacturer’s declaration – electromagnetic emissions
The Hemo-Force IPCS is intended for use in the electromagnetic environment specified below.
The customer or the user of the Hemo-Force IPCS should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic
environment – guidance
RF emissions
CISPR 11
Group 1
The
Hemo-Force IPCS
uses
RF energy only for its internal
function. Therefore, its RF emis
-
sions are very low and are not
likely to cause any interference
in nearby electronic equipment
RF emissions
CISPR 11
Class B
The
Hemo-Force IPCS
is
suitable for use in all establish-
ments, including domestic
establishments and those
directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
P
P
P