Maintenance
29
When the battery always gives a low battery indication or the battery is not installed correctly, a
battery check is needed.
6.3.2
Remove an internal battery.
Turn the power off and disconnect the power cord.
Use a screwdriver to loosen the battery cover fixing screws at the bottom of the pump.
Remove the battery cover.
Disconnect the battery cable connector.
Remove the battery.
6.3.3
Install the internal battery.
Insert the connector of the battery cable into the battery.
Insert the new battery into the battery compartment.
Attach the battery cover.
Use a screwdriver to tighten the screws securing the battery cover.
CAUTION:
Remove the battery if the infusion pump is not likely to be used for some time
WARNING:
l The battery’s replacement must be done by specialist who has been trained to finish such operation.
Otherwise there will be a risk of danger.
l Please strictly follow the instruction to replace the battery, and the battery should be provided by the
manufacturer. Otherwise there will be a risk of danger.
Do not disassemble, short circuit or throw the battery into fire in case of the danger caused by linkage or
explosion.