Calibration
39
patient side occlusion pressure. Various data of the user-defined brands is left blank when the pump is delivered.
If a user-defined IV set brand is not calibrated, users cannot use it.
a)
Method 1:
1
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, choose
Cali. Consum.
>
New
Consum.
to access the new consumables editing screen, and specify
Brand
,
Model
, and
Size
(
Brand
and
Size
are mandatory).
2
After the setting, tap
to access the
New Consum.
screen.
3
Manually input the rates of the 18 points of a known IV set, fluid side occlusion pressure, and level 11
patient side occlusion pressure.