Calibration
34
8
Calibration
8.1
Touch Screen Calibration
1
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, and choose
Device Cali.
>
Screen
Cali.
to access the touch screen calibration screen. Perform calibration based on the hint dot until the
prompt "
Calibration success
" is displayed, and tap
Confirm
to finish calibration.
8.2
Door Potentiometer Calibration
1
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, and choose
Device Cali.
>
Sensor
Cali.
>
Door Sensor Cali.
2
Select the motor rotation direction (
Clockwise
: consistent with the infusion direction), set the rotation
speed, and tap
Start
. The sliding block starts to slide, the mark line (crosshair) on the anti-free-flow clamp
is aligned with the mark lines on the sliding block respectively. Tap
Calibrate
at Calibrate 1 to Calibrate 4
respectively, as shown in the following figure.
Notes:
Calibrate 1: To calibrate the door opening status. If the calibration is not well performed, the door cannot
open or cannot open smoothly.
Calibrate 2 and 3: To calibrate the clamp status. If the calibration is not well performed, the anti-free-flow
clamp may be loose and liquid may leak.
Calibrate 4: To calibrate the occlusion pressure status. If the calibration is not well performed, the
monitoring status of fluid/patient side occlusion pressure may be incorrect, affecting the sensitivity of
triggering an occlusion alarm.