Calibration
36
(Picture of real product)
8.3
Pressure Sensor Calibration
1
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, choose
Device Cali.
>
Sensor
Cali.
>
Door Sensor Cali.
>
Pressure Sensor Cali.
, and input the k, b, upk, and upb values. The labels of
these values are affixed on the middle plate (or the pressure sensor).
8.4
Upper/Lower Occlusion Levitate Value Calibration
1
Open the pump door. The upper occlusion pressure sensor is in suspended state, and the upper occlusion
levitate value is displayed. Tap
Calibrate
to obtain the calibration value. The calibration is finished. Note:
If the levitate value is inaccurate, the zero point of the occlusion pressure may be affected, thereby affecting
the sensitivity of triggering an occlusion alarm.