Calibration
38
6
Cancel the calibration mode.
Note: Ensure that the section of the infusion tube to be installed onto the infusion pump has never been
extruded.
8.6
Patient Side Occlusion Pressure Calibration
1
Connect the pressure gage to a 2 Bar pressure sensor, perform relevant settings, and set the unit to
mmHg.
2
Connect the pressure gage to the IV set.
3
Install the infusion tube onto the infusion pump.
4
Rotate the three-way valve to ensure that the test IV set, backup syringe, and pressure gage are
connected and unblocked.
5
Perform purge until all air bubbles are removed from the infusion tube.
6
Rotate the three-way valve to ensure that only the test IV set and pressure gage are connected and
unblocked.
7
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, choose
Brand Maintenance
>
Cali. Consum.
>
Brand
, and select a corresponding IV set brand based on the IV set used.
8
Turn to the next page, that is, the
Down Press Cali.
page.
9
Wait for 5 minutes.
10
Set the pressure calibration speed to 0.06 r/s, and tap
Start
.
11
Observe the reading on the pressure gage. When the reading reaches the value (unit: mmHg) listed in
the following table, click
Read
in a corresponding line to refresh the pressure value.
Level
1
press
ure
Level
2
press
ure
Level
3
press
ure
Level
4
press
ure
Level
5
press
ure
Level
6
press
ure
Level
7
press
ure
Level
8
press
ure
Level
9
press
ure
Level
10
press
ure
Level
11
press
ure
Level
12
press
ure
150
225
300
375
450
525
600
675
750
825
900
975
12
Press
HOME
to return to the initial page.
8.7
Fluid Side Occlusion Pressure Calibration
1
Install the infusion tube onto the infusion pump, and block the input end.
2
Choose
HOME
>
Local Set
>
Maintenance
, input the password
1666
, choose
Brand Maintenance
>
Cali. Consum.
>
Brand
, and select a corresponding IV set brand based on the IV set used.
3
Access the
Down Press Cali.
page.
4
Set the pressure calibration speed to 10 r/s, and tap
Start
.
5
When the infusion tube is totally compressed, tap
Read
to refresh the value.
6
Press
HOME
to return to the initial page.
8.8
Infusion Pump Accuracy Calibration
Infusion set calibration purpose:
IV set calibration aims to increase the infusion accuracy of the IV set brands not built in the pump. The
parameters to be calibrated include the rates of the 18 points of the IV set, fluid side occlusion pressure, and
Use a slot type
screwdriver to
adjust the
peristalsis pump
position.