System Overview
Indicators
Power/working indicator: yellow or green
Alarm indicator: green, yellow, or red
Key backlight
Interfaces
Micro USB 2.0: enables connection to the external drop sensor.
USB 3.0: enables connection to a USB flash drive for software and
drug library upgrade.
USB2.0 barcode scanner/nurse caller/cabinet communication interfaces
RJ45 network interface: 10/100 Mbps adaptive Ethernet
WiFi network interface: 802.11-b/g/n, enables communication with the
infusion central monitoring system.
Infusion Rate
0.10 - 1200.0ml/h or 0.03-400d/min (20d/ml IV set)
0.10 - 400.0ml/h or 0.1-400d/min (60d/ml IV set)
Minimum Increment
of Infusion Rate
0.10-99.99ml/h (minimum increment: 0.01ml/h)
100-999.9ml/h (minimum increment: 0.1ml/h)
1000-1200ml/h (minimum increment: 1ml/h)
VTBI
0.10 - 9999.99ml (minimum increment: 0.01 ml)
Total Volume Display
0.01 - 9999.99ml (minimum increment: 0.01 ml)
Infusion Time
00:00:01 - 99:59:59 (minimum increment: 1s)
Bolus Rate
0.10 - 1200.0ml/h (20d/ml IV set)
0.10 - 400.0ml/h (60d/ml IV set)
BOLUS VTBI
0.10 - 50ml (minimum increment: 0.01 ml)
Anti-Bolus
Anti-bolus function, unintended bolus ≤ 0.2 ml
KVO Rate
0.1 - 5.0ml/h (minimum increment: 0.01 ml/h)
Purge Rate
1200.0ml/h (20d/ml IV set)
400.0ml/h (60d/ml IV set)
Infusion Accuracy
Non-dedicated IV set:
When infusion rate is smaller than 1 ml/h: infusion accuracy ≤ ±5%
When infusion rate is not smaller than 1 ml/h: infusion accuracy ≤ ±3%
Dedicated IV set: