Chapter 9: Circulation Mode
MBRAUN
systems work by the principle of gas circulation, i.e. the working
gas permanently circulates between the glove box and the H
2
O/O
2
gas
purification system. This process guarantees absolutely stable values of
purity and a cost-efficient processing.
Caution!
When operating the Glove Box system the circulation mode
should be activated, only in this case is
it
continuously purified to
values of < 1 ppm with regard to H
2
O/O
2
.
The circulation mode is SPS-controlled and is operated and displayed via the OP17 Operation Panel.
When used for a long period in the circulation mode
,
the purification unit gets exhaus
t
ed
,
resulting in a
drop of the purification performance leading to increasing H
2
O/O
2
values. For this reason, the purifier column
should be regenerated regularly
,
or at
the very least,
when there is a visible drop in performance. Refer to chapter
10 “Regeneration“.
The circulation mode should be deactivated while the regeneration procedure is running.
In systems with 2 purifier columns
,
circulation mode can run via one purifier column while the other purifier
column is undergoing regeneration.
Having observed all preceding chapters.
All connections are properly made.
All antechambers are closed.
The Glove Box system has been purged.
The system is activated.
No regeneration of the purifier column.
Glove
Box
Ultra
pure gas
Gas from
box
Gas
purification
General information
Prerequisites
Chapter 9
Circulation Mode
43
Summary of Contents for MB 120B-G
Page 6: ...OPERATING INSTRUCTIONS Chapter 1 General Information MBRAUN 1...
Page 10: ...OPERATING INSTRUCTIONS Chapter 2 Transport Site Selection Modification MBRAUN 2...
Page 13: ......
Page 14: ...OPERATING INSTRUCTIONS Chapter 3 Preparing the Connections MBRAUN 3...
Page 20: ...OPERATING INSTRUCTIONS Chapter 4 Installation MBRAUN 4...
Page 24: ...OPERATING INSTRUCTIONS Chapter 5 Activating and Deactivating the System MBRAUN 5...
Page 28: ...OPERATING INSTRUCTIONS Chapter 6 OP 17 Operation Panel MBRAUN 6...
Page 32: ...OPERATING INSTRUCTIONS Chapter 7 Purging the System MBRAUN 7...
Page 36: ...OPERATING INSTRUCTIONS Chapter 8 Pressure Control System MBRAUN 8...
Page 40: ...OPERATING INSTRUCTIONS Chapter 9 Circulation Mode MBRAUN 9...
Page 45: ......
Page 46: ...OPERATING INSTRUCTIONS Chapter 10 Regeneration MBRAUN 10...
Page 51: ......
Page 52: ...OPERATING INSTRUCTIONS Chapter 11 Settings and Display Patterns MBRAUN 11...
Page 59: ......
Page 60: ...OPERATING INSTRUCTIONS Chapter 12 Antechamber Operation MBRAUN 12...
Page 73: ......
Page 74: ...OPERATING INSTRUCTIONS Chapter 13 Dust Filters MBRAUN 13...
Page 77: ......
Page 78: ...OPERATING INSTRUCTIONS Chapter 14 Analyzers MBRAUN 14...
Page 90: ...OPERATING INSTRUCTIONS Chapter 15 Maintenance and Service MBRAUN 15...
Page 94: ...OPERATING INSTRUCTIONS Chapter 16 Spare Parts List MBRAUN 16...
Page 103: ......
Page 105: ......