Chapter 4: Installation
The initial installation of the system is recommended to be done by a competent
MBRAUN
technician only.
Caution!
When handling gases
,
potential for
accident
s
is permanently existent! The connection of the
system
should be carried out by competent and experienced person
n
el only!
MBRAUN
standard systems are not suited for using radioactive or toxic agents. In such a case,
special equipment components are required as well as special methods for the connections
and precautions have to be observed
.
M
oreover
,
special system components are provided not
described in this technical documentation. If necessary, the
MBRAUN
service department will
provide you with the pertinent information (e-mail: service@mbraun.
com
)!
Connecting the working gas
1. Connect the pressure reducing valve with the working gas source.
Observe the instructions for connection given by the manufacturer of the pressure reducing valve.
2. Make a supply line between the working gas source and the “Working gas INLET” system connection.
Observe chapter 3 “Preparing the connections”
3. The “Working gas INLET” system connection is marked by the exact value of the supply pressure.
Set pressure reducing valve to this value and open valve.
Caution!
Mind the exact pressure setting! Overpressure will damage the system - low pressure will cause
malfunction!
Connecting the regeneration gas
1. Connect the pressure reducing valve with the regeneration gas source.
Observe the pressure reducing valve manufacturer’s instructions for connections
2. Connect the working gas source with the “Regeneration Gas INLET” system connection using the supply pipe.
Observe chapter 3 “Preparing the connections”
3. The “Regeneration Gas INLET” system connection is marked by the exact value of the supply pressure.
Set pressure reducing valve to this value and open valve.
Caution!
Mind the exact pressure setting! Overpressure will damage the system - low pressure will cause
malfunction!
!
!
!
Safety instructions
Getting the system connected
Chapter 4
Installation
23
Summary of Contents for MB 120B-G
Page 6: ...OPERATING INSTRUCTIONS Chapter 1 General Information MBRAUN 1...
Page 10: ...OPERATING INSTRUCTIONS Chapter 2 Transport Site Selection Modification MBRAUN 2...
Page 13: ......
Page 14: ...OPERATING INSTRUCTIONS Chapter 3 Preparing the Connections MBRAUN 3...
Page 20: ...OPERATING INSTRUCTIONS Chapter 4 Installation MBRAUN 4...
Page 24: ...OPERATING INSTRUCTIONS Chapter 5 Activating and Deactivating the System MBRAUN 5...
Page 28: ...OPERATING INSTRUCTIONS Chapter 6 OP 17 Operation Panel MBRAUN 6...
Page 32: ...OPERATING INSTRUCTIONS Chapter 7 Purging the System MBRAUN 7...
Page 36: ...OPERATING INSTRUCTIONS Chapter 8 Pressure Control System MBRAUN 8...
Page 40: ...OPERATING INSTRUCTIONS Chapter 9 Circulation Mode MBRAUN 9...
Page 45: ......
Page 46: ...OPERATING INSTRUCTIONS Chapter 10 Regeneration MBRAUN 10...
Page 51: ......
Page 52: ...OPERATING INSTRUCTIONS Chapter 11 Settings and Display Patterns MBRAUN 11...
Page 59: ......
Page 60: ...OPERATING INSTRUCTIONS Chapter 12 Antechamber Operation MBRAUN 12...
Page 73: ......
Page 74: ...OPERATING INSTRUCTIONS Chapter 13 Dust Filters MBRAUN 13...
Page 77: ......
Page 78: ...OPERATING INSTRUCTIONS Chapter 14 Analyzers MBRAUN 14...
Page 90: ...OPERATING INSTRUCTIONS Chapter 15 Maintenance and Service MBRAUN 15...
Page 94: ...OPERATING INSTRUCTIONS Chapter 16 Spare Parts List MBRAUN 16...
Page 103: ......
Page 105: ......