4-27
MENU DESCRIPTIONS
www.maxcessintl.com
FIFE-500-XL
MI 2-267 1 D
The arrow buttons are used to jog the guide in either
direction. This can be used to move the web in and out of the sensor during
the calibration process.
Use the sensor select button to select the sensor input channel
before calibration. Pressing this button alternates between the S1 and S2
sensor input channels.
Press ACCEPT to save the results of the last sensor calibration and
return to the previous screen. This button is also required to save the results
of selecting a different calibration set.
Press CANCEL to exit without saving the results of the last calibration,
and return to the previous screen. Any changes to the sensor calibration set
is restored to its previous setting.
Calibration Procedure:
..
Select which set of calibration values will be used.
Select S1 or S2 sensor input.
Press the START CALIBRATION button. The bar graph indicator will be
rescaled to show the full signal range of –20mA to +20mA on the selected
sensor input. The sensor current will also be displayed under the bar graph
as shown is the example screen below.
Move the web material completely out of the sensor field of view and press
the CALIBRATE S1 UNCOVER button.
Move the web material into the sensor field of view until the entire
proportional band of the sensor is covered, and press the CALIBRATE S1
COVER button.
If the calibration was successful, the following message will appear.