7 Analog Control
43
R4G/R4GE series
7-3-1
Voltage Program Setting with External Voltage
The output voltage can be controlled with an external voltage. (Set SW1-8 to OFF.)
(Input voltage 0 to 10Vdc corresponds to the output voltage 0 to the rating.)
Check that the voltage setting mode is set to Remote mode.
(Refer to [7-2 Switching Local/Remote Mode (Voltage/Current Mode Setting].)
7-3-2
Voltage Program Setting with External Resistance
The output voltage can be controlled with an external resistance.
(Set SW1-7 to ON and SW1-8 to OFF.)
(Resistance value 0 to 10kΩ corresponds to the output voltage 0 to the rating.)
Check that the voltage setting mode is set to Remote mode.
(Refer to [7-2 Switching Local/Remote Mode (Voltage/Current Mode Setting].)
TB1
1
5
Internal circuit
+S
(common)
+
0 to 10Vdc
SW1
ON
1
2
3
4
5
6
7
8
TB1
1
5
Internal circuit
0 to approx.
10kΩ
SW1
ON
1
2
3
4
5
6
7
8
+
1mA
-
+S
(common)