6 Useful Features
31
R4G/R4GE series
5. In order to clear the multi-set function, press and hold the [SET] switch for approx. five seconds or
more.
The SET LED will stop blinking and the multi-set mode will be cleared.
*
The multi-set function returns to the state held just before the power off when the [POWER
ON/OFF] switch is turned on.
6-3
Startup Menu
In the Startup menu, setting the following various functions and initializing the unit are available.
Display
Item
Description
“dEF”: R4G series
“1”: R4GE series
Default
Initializes the unit to factory default.
“dly”
Delay trigger setting
Sets the delay trigger function. *
“unit”
Unit number setting
Sets the unit number. *
“PFP”: R4G series
“4”: R4GE series
Power failure protection function
setting
Sets the power failure protection function.
"LS"
Remote switch logic inverting
function setting
Sets the logic of the remote switch function. *
"Ld"
Interlock setting
Sets the interlock function. *
*
Only for the R4G series
6-3-1
Setting the Startup Menu
While holding the [SET] switch, turn on the [POWER ON/OFF] switch.
The display will switch to the "dEF" indication for the R4G series and the ”1” indication for the R4GE
series.
LOCK
OCP
SET
OVP/
CV
CC
V
A
OUTPUT
REMOTE
LOCK
OCP
SET
OVP/
CV
CC
V
A
OUTPUT
REMOTE
POWER ON
Pressing SET
“dEF”or “1”
Restore factory default
SET LED stops blinking.
Press and hold SET
LOCK
OCP
SET
OVP/
CV
CC
V
A
OUTPUT
REMOTE