Radical-7R Pulse CO-Oximeter Operator’s Manual
5-5
5
5-5
A l a r m s a n d M e s s a g e s
MESSAGE
POSSIBLE CAUSE(S)
RECOMMENDATION
INCOMPATIBLE AC
CABLE
Not a compatible cable.
Replace with a proper cable. Refer to
Section 8.
REPLACE AC
CABLE
Pulse CO-Oximeter
cannot identify the
connected cable or the
cable has failed.
Inoperative or faulty cable; Replace cable.
Refer to the Directions for Use of the cable
being used.
REPLACE AC
CABLE
Not a proper AC cable.
Replace with proper AC cable. Refer to
Section 8.
NO AC SENSOR
Sensor not fully inserted
into the connector.
May be an incorrect sensor, or a defective
sensor or cable. Insert sensor into
connector. Disconnect and reconnect
sensor. Refer to the instructions for the
sensor being used.
INCOMPATIBLE AC
SENSOR
Not a proper Masimo sensor.
Replace with a proper Masimo sensor. Refer
to Section 8.
REPLACE AC
SENSOR
Pulse CO-Oximeter
cannot identify the
connected AC sensor or
the sensor has failed.
Inoperative or faulty AC sensor; Replace
sensor. Refer to the Directions for Use of the
sensor being used.
AC SENSOR
Not a proper AC sensor.
Replace with proper AC sensor. Refer to Section 8.
AC SENSOR OFF
AC sensor not connected
to patient properly.
Sensor is damaged.
Properly reapply the AC sensor on the
patient and reconnect the sensor to the
instrument or patient cable.
If the sensor is damaged, replace the sensor.
S y s t e m M e s s a g e s ( c o n t i n u e d )