4-10
Radical-7R Pulse CO-Oximeter Operator’s Manual
4
Touch Key Control Buttons and Icons
On the Radical-7R Pulse CO-Oximeter display, four icons are shown on the right side of the LCD display.
These icons can be selected either by using the touch key control buttons or the touchscreen.
The touch key control buttons are the four dark grey control buttons to the right of the Handheld
display. To select an icon, press and release the touch key control button to the right of the icon,
or touch it on the touchscreen.
T r a d i t i o n a l U s e r I n t e r f a c e
CHANGE
LAYOUT
The change layout function is on button 1 when a menu is not up. It
allows the user to change the main layout between the one with the
SIQa and Respiration indicators.
MENU ACCESS
Press to enter the main menu.
SENSITIVITY
Press to toggle between the Normal, APOD and Maximum
Sensitivity modes. Use the Normal Sensitivity setting for typical
monitoring purposes. Use the APOD setting where there is a high
probability of the sensor becoming detached. Use the Maximum
Sensitivity setting for patients with low perfusion or when the low
perfusion message is displayed on the screen in APOD or normal
sensitivity mode. The default mode is APOD.
CAUTION:
WHEN USING THE MAXIMUM SENSITIVITY SETTING,
THE PERFORMANCE OF THE SENSOR OFF
DETECTION MAY BE COMPROMISED.
NOTE:
In “Custom” mode the instrument will remain in Normal
or APOD setting after a power cycle. Maximum Sensitivity will
automatically reset to Normal Sensitivity after a power cycle.
In “Neo” or “Adult” mode the instrument will reset the sensitivity to the
hospital specified setting (Normal or APOD) after a power cycle.
O p e r a t i o n