Radical-7R Pulse CO-Oximeter Operator’s Manual
3-3
3
S e t u p
S a t S h a r e S e t u p
The Radical-7R Pulse CO-Oximeter has been proven to be accurate during patient motion
and low perfusion conditions. Saturation and pulse rate values from the Radical-7R Pulse
CO-Oximeter may be displayed on a multiparameter monitor through the SatShare feature.
The SatShare feature provides an ideal, simulated waveform corresponding to the
measured saturation and pulse rate values determined by the Masimo SET technology.
This waveform may be used to display these values on multiparameter monitors through
the multiparameter oximetry sensor or input connector.
It is recommended that the Radical-7R Pulse CO-Oximeter is positioned close to the
multiparameter monitor with the Radical-7R Pulse CO-Oximeter screen visibly displaying
the plethysmographic waveform and the saturation and pulse rate measurements.
CAUTION:
SIMULTANEOUS USE OF SATSHARE AND SERIAL PORT IS NOT
SUPPORTED.
SATSHARE SETUP
1.
Select the SatShare cable that is appropriate for the multiparameter monitor that is
being connected. Check the Masimo web site at www.masimo.com for the latest list
of available SatShare cables and validated instruments.
2.
Connect the labeled end of the cable to the SatShare Cable Connector port on the
back of the Docking Station. Tighten the connector screws for a secure connection.
3.
Connect the other end of the SatShare cable either to the sensor connector of the
multiparameter monitor’s SpO
2
cable or directly to the SpO
2
connector on the monitor.
4. Verify that the Radical-7R Pulse CO-Oximeter recognizes the correct cable.
The name of the SatShare cable will be displayed on the LCD screen when the
SatShare mode is functional.
5. Set the multiparameter monitor’s high and low saturation and pulse rate alarm
limits as appropriate.
6. Set the multiparameter monitor’s averaging time to the lowest setting (i.e. fastest
response). The Radical-7R Pulse CO-Oximeter’s ideal waveform necessitates
the need for additional averaging by the monitor. If the multiparameter monitor’s
averaging time is not changed, the time to display physiological changes in
saturation on the monitor will be increased with SatShare. However, the delay can
be minimized by reducing the multiparameter monitor’s averaging time.
97
76
SpHb g/dl
17
7
10.0
SpOC