Radical-7R Pulse CO-Oximeter Operator’s Manual
4
S a t S h a r e O p e r a t i o n ( c o n t i n u e d )
■
While operating in the SatShare mode, the Radical-7R Pulse CO-Oximeter may
automatically disable the SatShare interface if the perfusion index drops below 0.1%
while the sensitivity is set to Max sensitivity. To enable the SatShare interface again,
set the Radical-7R Pulse CO-Oximeter to the Normal or APOD sensitivity mode,
increase the perfusion at the measurement site (by warming the patient or sensor
site), or move the sensor to a site with better perfusion.
CAUTIONS:
■
SATSHARE SIGNALS ARE IDEAL SIMULATED WAVEFORMS CORRESPONDING
TO THE CALCULATED SATURATION AND PULSE RATE VALUES AND DO
NOT CONTAIN ALL OF THE INFORMATION CONTAINED IN PHYSIOLOGICAL
WAVEFORMS. THE MULTIPARAMETER PATIENT MONITOR DECODES THESE
SIGNALS INTO SATURATION AND PULSE RATE VALUES.
■
DURING SATSHARE OPERATION, THE AUDIBLE ALARMS MAY BE MUTED
ON THE RADICAL-7R PULSE CO-OXIMETER. WHEN THE AUDIBLE ALARM
IS MUTED (INDICATED BY A BELL WITH A SLASH THROUGH IT) ON THE
RADICAL-7 PULSE CO-OXIMETER, USE THE MULTIPARAMETER MONITOR FOR
AUDIBLE ALARM INDICATION.
■
THE (RADICAL-7R) SpO
2
AND PULSE RATE AUDIBLE ALARMS MAY BE
DISABLED WHEN THE RADICAL-7R IS CONFIGURED WITH RAINBOW
PARAMETERS/MEASUREMENTS. REFER TO SECTION 4
ALARMS.
■
THE RADICAL-7R REVERTS TO THE INTERFACE ALARM "SpO
2
/BPM" “Yes”
SETTING DURING POWER INTERRUPTIONS OR WHEN THE SATSHARE
CONNECTION IS LOST OR THE DEVICE BECOMES SEPARATED FROM THE
DOCKING STATION. THIS ENSURES THAT THE RADICAL-7R PROVIDES
AUDIBLE ALARMS FOR SpO
2
AND PULSE RATE WHEN THE CONNECTION TO
THE INTERFACED SYSTEM BECOMES COMPROMISED.
■
DURING SATSHARE OPERATION DO NOT USE THE PLETHYSMOGRAPHIC
WAVEFORM DISPLAY ON THE MULTIPARAMETER MONITOR FOR DIAGNOSTIC
PURPOSES. INSTEAD, USE THE PLETHYSMOGRAPHIC WAVEFORM DISPLAYED
ON THE RADICAL-7R SCREEN.
■
TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY
EQUIPMENT TO THE SATSHARE CONNECTOR UNLESS THE RADICAL-7R IS
CONNECTED TO THE AC MAINS POWER SUPPLY.
■
ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED.
■
ONLY
SPO
2
AND PULSE RATE CAN BE DISPLAYED ON THE MULTIPARAMETER
MONITOR WITH SATSHARE.
To return from SatShare operation to normal standalone operation, simply disconnect
the SatShare cable from the patient monitor or disconnect the SatShare cable from the
SatShare connector on the back of the Radical-7R Pulse CO-Oximeter.