Radical-7R Pulse CO-Oximeter Operator’s Manual
2-9
2
S y s t e m D e s c r i p t i o n
STANDALONE BACK PANEL
1
SERIAL OUTPUT
CONNECTOR
Use the Serial Output Connector with a ferrite bead
installed to connect a serial instrument, including a serial
printer, a monitoring system or PC to the Radical-7R Pulse
CO-Oximeter. The data is provided in standard RS-232C
format. See Section 7,
Serial Interface Specifications
.
All external instrument connections to the Serial Output
Connector must be IEC-60950 compliant.
2
ANALOG OUTPUT/
NURSE CALL
CONNECTOR
Use the Nurse Call Connector with a ferrite bead installed
to interface with an analog output instrument, such as a
chart recorder or nurse call system. All external instrument
connections to the Analog Output / Nurse Call Connector
must be IEC-60950 compliant.
3
SATSHARE CABLE
CONNECTOR
Use the SatShare Cable Connector to connect a SatShare cable to
the SpO
2
input connector of a multiparameter patient monitor. All
external instrument connections to the SatShare Cable Connector
must be IEC-60601-1-1 compliant. SatShare cables are available to
interface with most major multiparameter patient monitors. Check the
label on the SatShare cable and the SatShare Directions for Use to
ensure that the correct cable is used for each type of patient monitor.
4
POWER ENTRY
MODULE
The power entry module contains the input connector for
AC power and two fuses. The AC input provides power to
the system from the AC line. Always connect the Pulse
CO-Oximeter to the mains power for continuous operation and/
or battery recharging.
NOTE:
Use the power cord as the means to disconnect the
instrument from the mains power supply.
5
EQUIPOTENTIAL
GROUND
CONNECTOR
Use the Equipotential Ground Connector for grounding.
1
2
3
4
5