Radical-7R Pulse CO-Oximeter Operator’s Manual
4-17
4
G e n e r a l
MENU ITEMS
DESCRIPTION
FASTSAT
Select
Yes
to activate the FastSat algorithm. In the 2 and 4 seconds
averaging mode, the FastSat algorithm is automatically enabled.
SMART TONE
Select
Yes
to activate the SmartTone function. This will allow the audible pulse to
continue to beep when the pleth graph shows signs of motion. Select
NO
to turn
off SmartTone.
PULSE TONE
Select YES to activate the Pulse Tone function. This will allow the audible
pulse to continue to beep upon each patient pulsation..
A b o u t
This displays the copyright and software versions of the Handheld and Docking Station.
MENU ITEMS
Copyright Date
Masimo Corporation
Masimo SET Software Version
Handheld Software Version
D-Station Software Version
C o n f i g u r a t i o n
MENU ITEMS
DESCRIPTION
Line Frequency
Set to 50 or 60Hz.
Set to match regional power line frequency to allow for
cancellation of noise introduced by fl uorescent lights and other
sources. Factory default is 60 Hz.
Rapid Desat Limit
Set to Off, -5 or -10.
When set to -5%, this feature overrides the Alarm Delay (audible
alarms silenced) and the instrument will alarm if SpO2 rapidly falls to
5% or 10% below the alarm limit. Factory default is -5%.
SpHb Cal
Set to Arterial or Venous.
This feature provides an SpHb (Arterial) or SpHbV (Venous)
value that displays on the main screen.
NOTE: The hemorheologic profi le of arterial and venous blood
samples can vary1. To accommodate this difference, the
Radical-7R provides the option of displaying a SpHb parameter
that is based on either Arterial or Venous SpHb laboratory
blood sample draw.
Changes to SpHb are included in output data. SpOC
calculation will always be based on the arterial SpHb.
O p e r a t i o n