ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
|
54
|
10 Technical Data
|
ROTAFLOW with ICU Kit
|
10.2
ROTAFLOW CONSOLE
10.2.1
POSSIBLE SETTINGS
Power supply
Stand-alone operation
Mains power supply
Line voltage
(factory-set)
100/115/230/240 V AC
Fuse
For 100/115 V: 2 A
For 230/240 V: 1 A
Frequency
50/60 Hz
Operation on heart-lung
machine
Power supply via heart-lung machine
Voltage
+24 V
Current
Max. 8 A
Batteries
Type
NiCd
Voltage
24 V
Capacity
5 Ah
Charge/discharge cycles
Min. 500
Shelf life
Max. 6 months
(under storage conditions)
Serial output interface
Transfer rate
9600 Baud
Protocol
8 N 1
Measurements and weight
Dimensions (length × width ×
height)
179 × 385 × 243 mm
Weight
Approx. 14.4 kg
Status display
Meaning
Possible values
SET MODE
Mode
ART, ART PULS, STAND AL, FREE
(
SET BRIG
Brightness of display
25%, 50%, 100%
SET FLIM
Lower flow limit
Off,
0.1 … 0.9 l/min (resolution 0.1 l/min),
1.0 … 5.0 l/min (resolution 0.5 l/min)
SET LLIM
Upper speed limit
Off,
500 … 4500 rpm (resolution 500 rpm)
SET HLIM
Upper speed limit
Off,
3000 … 5000 rpm (resolution 500 rpm)
SET CLAM
Not used: Automatic or manual tube clamp
Summary of Contents for ROTAFLOW
Page 1: ...USER S MANUAL ROTAFLOW WITH ICU KIT ROTAFLOW SYSTEM...
Page 57: ......