5 Safety
|
ROTAFLOW with ICU Kit
|
11
Fit the protective cap to the connection socket of the heart lung machine.
5.5
USING EMERGENCY DRIVE
A ROTAFLOW System can be easily replaced with a spare ROTAFLOW System.
In emergencies, the ROTAFLOW Centrifugal Pump can be operated manually with
the ROTAFLOW Emergency Drive.
Bubble and level monitoring in “STAND AL” mode
Monitoring is performed by the internal bubble sensor of the ROTAFLOW Drive
and a level sensor, which is connected to the ROTAFLOW Console.
In “STAND AL” mode the pressure is not monitored.
1
Close the arterial and the venous
tube clamp.
2
Remove the ROTAFLOW Centrifugal
Pump from the ROTAFLOW Drive.
Insert it in the ROTAFLOW
Emergency Drive.
3
Lift the handle and turn clockwise.
The speed display shows the speed
from 1500 rpm to 5000 rpm in steps
of 500 rpm.
Above 5000 rpm, all LEDs are lit.
4
When the speed is sufficiently high,
open both tube clamps.
Note
Turning counterclockwise is not possible.