ROTAFLOW with ICU Kit
|
4 Operation
|
31
|
ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
4
OPERATION
4.1
MODES
The relevant mode must be selected for the ROTAFLOW System in each case.
When selecting the mode, take note of
4.2 “Selecting settings”, page 33 and
5.1 “Modes “ART” and “ART PULS”: Communication error with heart-lung
machine”, page 39.
WARNING!
Always have a ROTAFLOW Emergency Drive easily accessible and ready for
use.
Set the patient-specific lower flow limit (FLIM) correctly (
settings”, page 33). If the flow is too low, this may be dangerous and harm the
patient.
Before starting the application, check the items listed in
Only ever start the application when the battery of the ROTAFLOW Console is
fully charged.
WARNING! "FREE" mode: No interventions
In “FREE” mode, no intervention is made by the monitoring system.
Do not use “FREE” mode during an operation, except in an emergency.
Display in “SET
MODE” menu
Shown in status
display
Meaning
ART
ART
Only with operation on a heart-lung machine:
Arterial pump for continuous pumping
Interventions using heart-lung machine (
4.7 “Modes “ART” and “ART PULS”:
Interventions”, page 38).
See also the User’s Manual for the heart-lung
machine.
ART PULS
A-P
Only with operation on a heart-lung machine:
Arterial pump with a pulsatile pumping option
Interventions using heart-lung machine (
4.7 “Modes “ART” and “ART PULS”:
Interventions”, page 38).
See also the User’s Manual for the heart-lung
machine.