ROTAFLOW with ICU Kit
|
10 Technical Data
|
53
|
ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
10
TECHNICAL DATA
10.1
GENERAL
10.1.1
SPEED
10.1.2
FLOW
10.1.3
AMBIENT CONDITIONS
Range
0 … 5000 rpm
Display resolution
1 rpm
Display accuracy
±20 rpm
Range
0 … 9.99 l/min
Display resolution
0.01 l/min
Display accuracy
From 0 to 1 l/min:
0.07 l/min + offset drift
From 1 to 9.99 l/min:
7% of current value + offset drift
Offset drift
Max. 0.05 l/min
Maximum flow deviation with an
error
+20%
Operation
Storage
Transport
Temperature
+10 … +40ºC
-18 … +45ºC
-18 … +55ºC
Relative humidity
(non-condensing)
15 … 95%
10 … 96%
10 … 96%
Air pressure
66 … 106 kPa
66 … 106 kPa
66 … 106 kPa
Note
If storage and transportation exceed 15 weeks, comply with the ambient