6 Cleaning and disinfection after each use
|
45
|
ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
6
CLEANING AND DISINFECTION AFTER EACH USE
Clean the surface of the whole system to remove any spilt blood.
If necessary, use a scrubbing sponge suitable for stainless steel and aluminum
surfaces.
Standard liquid household cleaners can generally be used to clean the foils on the
front panel. In case of doubt, please consult a MAQUET representative.
Open the lock on the ROTAFLOW Drive and remove all the ultrasonic contact
cream with a lint-free cloth.
ATTENTION! Risk of damage to equipment
Make sure that no liquids enter the ventilation slit of the ROTAFLOW Drive.
Never flush the system with large amounts of water.
Do not use chemical solvents (e.g., alcohol, ether or acetone).
Do not spill anesthetics (e.g., Foram (isoflurane)).