ROTAFLOW with ICU Kit
|
3 Connection and Installation
|
25
|
ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
3.3
INSTALL ROTAFLOW DRIVE
1
Use the wing bolt to secure the ROTAFLOW Drive to a stable pole (diameter
25 mm … 40 mm) or to the holder for the ROTAFLOW Drive on the
ROTAFLOW Console.
2
Make sure that the on/off switch of the ROTAFLOW Console is switched off.
3
Connect the cable of the ROTAFLOW Drive to the “RFD Master” socket.
3.4
INSTALL ROTAFLOW EMERGENCY DRIVE
In emergencies operation can be maintained manually using a ROTAFLOW
Emergency Drive.
Use the wing bolt to secure the ROTAFLOW Emergency Drive to a stable pole.
Secure on the same side as the ROTAFLOW Drive.
WARNING!
In transport applications, all components must be attached to the holders
provided for this purpose and securely fastened (
Ö
Instructions supplied with
holders for ELS System).
ATTENTION!
Switch off the ROTAFLOW Console before connecting or disconnecting the
cable of the ROTAFLOW Drive.
Make sure that the ventilation slit of the ROTAFLOW Drive is not blocked
during operation.
Make sure that no liquids enter the ventilation slit of the ROTAFLOW Drive.
Do not tip the ROTAFLOW Drive at an angle of more
than 90 degrees and do not turn upside down, as liquids
may enter the ventilation slit.
When using the ROTAFLOW Drive on the
ROTAFLOW Console, secure the
ROTAFLOW Drive in a position above the
ROTAFLOW Console to ensure greater
stability.