ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
|
24
|
3 Connection and Installation
|
ROTAFLOW with ICU Kit
|
1
Please make sure that the pump
switch on the console of the heart
lung machine is switched off.
2
Remove the protective cap from the
connection socket of the heart lung
machine.
3
Open the protective plug cap of the
ROTAFLOW Console and lock in the
securing mechanism.
4
Set the ROTAFLOW Console down
carefully on the console of the heart-
lung machine. The holding pins must
be inserted in the housing of the
ROTAFLOW Console.
5
Switch on the pump switch on the
console of the heart-lung machine.
Summary of Contents for ROTAFLOW