9 Messages
|
ROTAFLOW with ICU Kit
|
9.2.3
IN FLOW DISPLAY
RUN AWAY
Yes
Yes
Speed is more than 20% faster than the set value
(above 1000 rpm).
Turn the flow regulator to zero to clear the error.
With continuous pumping, the reaction time is
approx. 5 seconds. With pulsatile pumping, the
reaction time is approx. 15 seconds
(1.5 seconds above 6000 rpm).
VALVE?
-
-
Confirmation required that suitable systems or
methods are available to prevent a reverse flow
(
3.7 “Switch on ROTAFLOW Console, Self-test”,
Message
Acoustic
alarm
Pump
stop
Cause/Error source
-
∗ ∗
-
Flow meter not connected.
– – – –
Yes
No
Invalid flow value
⇑⇑⇑⇑
–
–
Flow > 10 l/min
The ROTAFLOW Console stops the pump at
12 l/min and an acoustic alarm is emitted.
⇓⇓⇓⇓
Yes
No
Flow < lower flow limit (FLIM)
????
Yes
Yes
Flow meter
DEL
Yes
Yes
Delay line of flow meter
ERR
Error: Flashes together with a further message.
RAM
Yes
Yes
Flow meter memory
SIG!
Yes
No
Error in flow measurement
Replacing ultrasonic contact cream (
41)
During longer procedures, it is possible that the
ultrasonic contact cream might dry out, which can
impair the results of the flow meter. The
ROTAFLOW Centrifugal Pump continues to
function nevertheless.
STAT
Yes
Yes
Flow meter status
TEMP
Yes
Yes
Flow meter temperature
TXRX
Yes
Yes
Transmission receiver of flow meter.
Message
Acoustic
alarm
Pump
stop
Cause/Error source