![Maico MB 11 BERAphone Operation Manual Download Page 77](http://html1.mh-extra.com/html/maico/mb-11-beraphone/mb-11-beraphone_operation-manual_3713323077.webp)
Operation Manual MB 11 BERAphone
®
/Classic
8100496 Rev. 6
75
23/10/2019
ABR
Elektrode
impedance
measurement
Measurement
current
200 nA
Waveform
Rectangular
Acceptable
Impedance
Range
0.25-
15 kΩ Vertex,
0.25-
15 kΩ Nape (Classic)
0.25-
10 kΩ Mastoid (BERAphone
®
)
Accuracy of
Measurement
Sensitivity
> 99.6 %
Specificity
97.9 %
1
TRANSDUCER
Radioear IP30
Type
ABR insert phones (50
Ω)
Versions
With EarCup
TM
adapter or eartip adaper
Supported tests
Binaural or monaural ABR
Max. input voltage
5.0 V RMS
THD
< 2% (125 Hz
– 4 kHz)
Cable length
37 cm / 14.5 in
Tube length
25 cm / 9.8 in
Tube colors
Red (right ear) and blue (left ear)
Weight (incl. cables)
53 g / 1.87 oz
BERAPHONE
®
Preamplifier
Channels
One
Gain
69.6 dB
Frequency response
35 Hz to 4.3 kHz
Noise
<25 nV/√Hz
CMR Ratio
> 120 dB at 80 Hz
Input impedance
1100 MΩ / 32 pF, 98 MΩ at 50 Hz
Power supply
Isolated, from USB box
Speaker
Integrated
Dynamic, 8 Ω
Electrodes
3pcs. with gel protectors Stainless-steel, reusable, rotatable vertex
electrode
Properties
Weight
285 g/ 10.1 oz
Dimensions
160 mm x 87 mm x 60 mm /
1
Cebulla, M., & Shehata-Dieler, W. (2012). ABR-based newborn hearing screening with
MB11 BERAphone
®
using an optimized chirp for acoustical stimulation.
International Journal
of Pediatric Otorhinolaryngology
,
76
(4), 536-543
.
Summary of Contents for MB 11 BERAphone
Page 1: ...Operation Manual MB 11 BERAphone Classic ...
Page 2: ......