Operation Manual MB 11 BERAphone
®
/Classic
8100496 Rev. 6
41
23/10/2019
New Patient
Click on the button
New
or
select
New
from the menu
File
to clear all patient
information fields, if they contain information on a previously tested patient.
To store the patient and measurements in the database or to print results it is required
to enter the
Last name
and
First name
, date of birth (
Born
), gender (
male
/
female
)
and Patient ID (
Pat. ID
).
The current PC date is displayed as the default value for the date of birth. Use the
calendar control to modify the date of birth. Select the gender male or female by the
radio button. Before a patient is stored in the database, the patient information can be
edited or text deleted by pressing the
Backspace
button. As soon as you start a
measurement or add the patient to the Today’s List, it gets saved in the database.
Once the patient data is saved, you cannot edit or delete the data.
Make sure the entered Patient ID is unique. If a Patient ID is entered which is already
stored in the database, a warning is displayed. Either change the Patient ID or load the
patient with this Patient ID from the database via
Search
(see section 5.1.6).
Search Patient
Click on the button
Search
to show a list of all stored patients in the
Load from
Database
dialog (see Section 5.6.2).
Select a patient by double-clicking on the name with the mouse or highlight the desired
name and select the
OK
button.
Database filters are available in the
Load from Database
dialog to find patients more easily.
Enter a
Pat. ID
or
Last Name
to filter the list by your entry. Select a date range or result
(
PASS
,
REFER
or
ABORT
) to find patients with measurements within the
Range of Date
or the corresponding
Test Result
. Select a screener name at
Screener
instead of
All
Screeners
to show all patients with tests performed by the selected screener. Only screener
names with stored tests in the database are available
in the “screener” list in this dialog.
Preparing for Testing
Testing environment
5.2.1.1 General
Keep in mind the indications and contraindications of use
given in sections 1.2 and 1.3.
Hearing screening is most successfully and efficiently performed on a quiet, sleeping
baby. If the baby is awake but quiet or sucking intermittently, testing is possible though
the test time may be affected. If the baby is crying, moving or sucking vigorously and
constantly then the test will be prolonged and the chance of a
REFER
result will be
increased. In this case it would be best to terminate the screening and return when the
baby is sleeping.
WARNING