Operation Manual MB 11 BERAphone
®
/Classic
8100496 Rev. 6
46
23/10/2019
NOTE
: Skin preparation products vary in terms of abrasiveness. Be sure to follow
the instructions on the product to avoid damage to the skin.
Preparing the skin helps to achieve good contact (i.e. low impedance) between the skin and
the electrode. After cleaning, remove any residue of the skin prep product so that the skin
is dry. This will help to ensure good adhesion of the disposable electrode to the skin.
Risk of strangulation.
Keep cables away from baby’s neck.
Avoid contact between the unused electrodes and
any other conductive parts.
Figure 24
Connect the white, black and yellow pinch clip electrode
lead wires to a snap electrode (Figure 24).
Figure 25
Figure 26
Peel the electrodes from the backing card and place
them on the electrode positions following this color
scheme. A graphic near the electrode jacks on the
preamplifier illustrates proper placement for the nape
montage as a reminder (Figure 25, see also Table 9 for
nape and mastoid montage).
Table 9 Nape and Mastoid Montage
Nape Montage
(recommended)
Mastoid Montage
(alternate)
Yellow
Forehead
Forehead
Black
Cheek (either side) Right mastoid
White
Nape of the neck
Left mastoid
Press gently around the entire surface of each electrode
to help secure its adhesive to the skin (Figure 26).
WARNING
CAUTION